Covidien, LP: Medical Device Recall in 2024 - (Recall #: Z-1457-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Product Classification:
Class II
Date Initiated: February 14, 2024
Date Posted: April 10, 2024
Recall Number: Z-1457-2024
Event ID: 94128
Reason for Recall:
Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
Status: Ongoing
Product Quantity: 30 Units
Code Information:
Model Number: SIGSDS30CTVT; UDI/DI: 10884521741881; Lot number: N3L1909UY;
Distribution Pattern:
Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.
Voluntary or Mandated:
Voluntary: Firm initiated