Product Description:

Prisma M 100 PRE set. Product number 103657. Each set is packaged in a plastic pouch, 4 pouches per carton box

Product Classification:

Class II

Date Initiated: July 18, 2014
Date Posted: August 20, 2014
Recall Number: Z-2197-2014
Event ID: 68788
Reason for Recall:

Gambro initiated a field action on the Prisma and Prismaflex sets due to breakages of connectors resulting in leakage of liquid or blood.

Status: Terminated
Product Quantity: 401,451
Code Information:

Prisma M 100 PRE set. Product number 103657. Batch number 12A0302G to 14B2102. Expiry 01/2014 to 02/2016

Distribution Pattern:

Worldwide Distribution.

Voluntary or Mandated: