Medica Corporation: Medical Device Recall in 2018 - (Recall #: Z-1149-2018)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2018.

Data Source: FDA.
Product Description:

Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.

Product Classification:

Class II

Date Initiated: November 21, 2017
Date Posted: March 28, 2018
Recall Number: Z-1149-2018
Event ID: 79264
Reason for Recall:

The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.

Status: Terminated
Product Quantity: 392
Code Information:

Lots 16279, 17045 and 17215

Distribution Pattern:

Worldwide and US Distribution.

Voluntary or Mandated:

Voluntary: Firm initiated