3M Company - Health Care Business: Medical Device Recall in 2013 - (Recall #: Z-2242-2013)
See the recall detail below. You can also see other recalls from the same firm in 2013.
3M Scotchcast, Wet or Dry Cast Padding, Catalog Numbers WDP2, WDP3, WDP4, and WDP6. Intended for use in constructing casts for either wet or dry immobilization.
Class II
3M is conducting a voluntary product recall of selected lots of 3M" Scotchcast" Wet or Dry Cast Padding. Upon investigation of a health care provider feedback, and confirmed with internal testing, it was found that the cast padding material does not consistently repel water. Lack of repellency may result in prolonged cast dry times after exposure to wet conditions. There have been no repo
Catalog # - Lot # WDP2 - 792920, 793090, 793640 & 796078. WDP3 - 792921, 793098, 793638, 796077 & 798487. WDP4 - 792922, 793116, 793637, 796079 & 798488. WDP6 792923, 793636, 798489, 799235 & 799236
Worldwide Distribution - USA including AL, AZ, AR, CA, CO,CT, FL, GA, HI, IL, IN,KS,KY, LA, MD, MA, MI, MN, MO, NE, NV, NJ, NY, NC, OH, OR,PA, TN, TX, UT,VA, WA, and WI. Internationally to GERMANY, CANADA, BELGIUM, NETHERLANDS, HONG KONG, NEW ZEALAND, AUSTRALIA, INDIA, KOREA, SINGAPORE AND UNITED ARAB EMIRATES, JAPAN, JAMAICA..
Voluntary: Firm initiated