3M Company - Health Care Business: Medical Device Recall in 2023 - (Recall #: Z-2637-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

3M Attest Super Rapid Readout Biological Indicator, REF 1492V; used to qualify or monitor dynamic-air-removal steam sterilization cycles

Product Classification:

Class II

Date Initiated: August 22, 2023
Date Posted: October 4, 2023
Recall Number: Z-2637-2023
Event ID: 92915
Reason for Recall:

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

Status: Ongoing
Product Quantity: 79500 devices
Code Information:

UDI/DI: 50707387762969 (shipper), 30707387762965 (carton); Lot codes: 33P3X3, exp. 12/29/2024; 33P4NH, exp. 01/01/2025

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.

Voluntary or Mandated:

Voluntary: Firm initiated