3M Attest Super Rapid Readout Steam Challenge Pack, REF 1496V; designed to qualify and monitor dynamic-air removal steam sterilization processes

Product Classification:

Class II

Date Initiated: August 22, 2023
Date Posted: October 4, 2023
Recall Number: Z-2638-2023
Event ID: 92915
Reason for Recall:

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

Status: Ongoing
Product Quantity: 33504 devices
Code Information:

UDI/DI: 50707387762983( (shipper); Lot codes: 33P4NJ, exp. 01/01/2025; 33P56T, exp. 01/02/2025

Distribution Pattern:

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.

Voluntary or Mandated:

Voluntary: Firm initiated