3M Company - Health Care Business: Medical Device Recall in 2023 - (Recall #: Z-2640-2023)
See the recall detail below. You can also see other recalls from the same firm in 2023.
3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482VF; designed to qualify and monitor dynamic-air removal steam sterilization processes
Class II
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.
UDI/DI: 50707387787283( (shipper); Lot Codes: 41482VF, exp. 33P4ND; 33P4YN, exp. 01/02/2025
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
Voluntary: Firm initiated