3M Healthcare Business: Medical Device Recall in 2022 - (Recall #: Z-0258-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2022.

Data Source: FDA.
Product Description:

3M Steri-Drape, Medium Drape with Adhesive Aperture, REF 1030NS, general surgery drape

Product Classification:

Class II

Date Initiated: October 6, 2022
Date Posted: November 30, 2022
Recall Number: Z-0258-2023
Event ID: 90989
Reason for Recall:

During a recent investigation, 3M confirmed the liner on the adhesive component of the affected drapes is difficult to remove without damaging the product and may render the product unusable. Additionally, 3M has observed an increase in reported adhesive related skin injuries for these affected lots.

Status: Ongoing
Product Quantity: 43,200 units
Code Information:

UDI/DI (01)50707387029352, Lot numbers: 33JPRN, 33JPWY, 33JR98, 33KLEY, 33KLME

Distribution Pattern:

Worldwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated