AB Sciex, LLC: Medical Device Recall in 2023 - (Recall #: Z-0037-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

API3200MD Mass Spectrometer, Part Number 4466230

Product Classification:

Class II

Date Initiated: August 30, 2023
Date Posted: October 11, 2023
Recall Number: Z-0037-2024
Event ID: 92810
Reason for Recall:

In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.

Status: Ongoing
Product Quantity: 3
Code Information:

Serial Numbers: EQ230011907, EQ230021911, EQ230001901

Distribution Pattern:

US Nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated