Abbott Diabetes Care, Inc.: Medical Device Recall in 2023 - (Recall #: Z-1273-2023)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

FreeStyle Libre Reader, REF: 71525-01, 71701-01, 71739-01 a component of the FreeStyle Libre Flash Glucose Monitoring System

Product Classification:

Class I

Date Initiated: February 13, 2023
Date Posted: April 12, 2023
Recall Number: Z-1273-2023
Event ID: 91756
Reason for Recall:

Lithium-ion batteries in glucose monitoring system readers may swell, overheat, or pose fire hazard. Only use the charging cable/power adapter included with readers. Do not use reader if it is damaged, cracked, swelling, too hot to hold, it will no longer hold charge, or it fails Reader Test. Do not use other chargers. User Manual update includes instructions on reader safe storage, charging, use.

Status: Ongoing
Product Quantity: 336,085
Code Information:

REF/UDI-DI/Lot: 71525-01/357599000219/02Y702D, 01Y699D, 02Y699D, 01Y698D, 02Y697D, 04Y690D, 03Y690D, 01Y683D, 02Y683D, 01Y674D, 02Y672D, 01Y669D, 02Y666D, 01Y659D, 02Y658D, 02Y656D, 04Y655D, 01Y655D, 01Y652D, 02Y653D, 03Y653D, 03Y651D, 01Y641D, 02Y638D, 01Y637D, 02Y635D, 01Y629D, 03Y628D, 01Y628D, 02Y628D, 02Y621D, 02Y622D, 02Y620D, 02Y608D, 04Y608D, 01Y608D, 01Y603D, 03Y599D, 04Y600D, 02Y596D, 06Y596D, 03Y596D, 07Y547D, 07Y546D, 08Y546D, 05Y540D, 06Y538D, 05Y537D, 03Y536D, 04Y523D, 02Y522D, 02Y519D, 02Y512D, 01Y512D, 01Y511D, 02Y510D, 01Y505D, 02Y504D, 04Y355D, 02Y355D, 01Y352D, 01Y349D, 03Y348D, 01Y348D, 02Y348D, 01Y339D, 01Y335D, 02Y334D, 02Y335D, 03Y314D, 01Y313D, 02Y312D, 01Y312D, 02Y311D, 01Y311D, 02Y310D, 04Y310D, 02Y307D, 03Y307D, 01Y304D, 02Y303D, 02Y304D; 71701-01/357599000226/01Y702D, 01Y694D, 03Y693D, 02Y686D, 03Y683D, 02Y574D, 01Y563D, 02Y563D, 01Y564D, 03Y562D, 01Y552D, 02Y552D, 06Y550D, 05Y539D, 05Y538D, 06Y537D, 01Y536D, 01Y532D, 02Y531D, 03Y522D, 02Y518D, 02Y511D, 01Y508D, 02Y505D, 04Y356D, 01Y356D, 02Y356D, 01Y321D, 01Y320D, 02Y320D, 01Y317D, 02Y317D; 71739-01/357599000240/03Y754D, 03Y734D, 01Y705D, 01Y671D

Distribution Pattern:

U.S. Nationwide.

Voluntary or Mandated:

Voluntary: Firm initiated