Abbott GmbH: Medical Device Recall in 2024 - (Recall #: Z-1479-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
ARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35
Class II
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).
a) 1L79-25, 00380740001414, Lot Numbers: 52011BE00, 54140BE00, 56375BE00, 58163BE00, 59194BE00; b) 1L79-35, 00380740001421, Lot Numbers: 52015BE00, 54144BE00, 56379BE00, 58167BE00, 59198BE00
US Nationwide distribution.
Voluntary: Firm initiated