Abbott Ireland Diagnostics Division: Medical Device Recall in 2019 - (Recall #: Z-0665-2019)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2019.
Data Source: FDA.
Product Description:
Alinity i Free T4 Reagent Kit, List Number 07P7030
Product Classification:
Class II
Date Initiated: November 21, 2018
Date Posted: January 2, 2019
Recall Number: Z-0665-2019
Event ID: 81632
Reason for Recall:
Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.
Status: Terminated
Product Quantity: 294 units
Code Information:
UDI/Lot Numbers (01)00380740121822 (17)190727(10)90111UI00 / 90111UI00 (01)00380740121822 (17)190911(10)92388UI00 / 92388UI00
Distribution Pattern:
AR, FL, GA, MD, OK, VA
Voluntary or Mandated:
Voluntary: Firm initiated