Abbott Laboratories: Medical Device Recall in 2021 - (Recall #: Z-2185-2021)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2021.

Data Source: FDA.
Product Description:

REF 8L92-22, Abbott Activated Alanine aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161859

Product Classification:

Class II

Date Initiated: June 10, 2021
Date Posted: August 11, 2021
Recall Number: Z-2185-2021
Event ID: 88123
Reason for Recall:

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

Status: Ongoing
Product Quantity: 1,603 units
Code Information:

Lot Numbers / Expiration Date 02954UN20 / 29AUG2021 37921UN20 / 01DEC2021 81823UN21 / 16FEB2022

Distribution Pattern:

Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.

Voluntary or Mandated:

Voluntary: Firm initiated