Abbott Medical Optics, Inc.: Medical Device Recall in 2013 - (Recall #: Z-0047-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

STAR ActiveTrak Excimer Laser System, Model: STAR S3, Catalog Number: 0030-2450, Product is manufactured and distributed by AMO Manufacturing USA, LLC, Milpitas, CA The STAR Excimer Laser Systems are indicated for laser assisted in situ keratomileusis (LASIK), photorefractive keratectomy (PRK), and phototherapeutic keratectomy (PTK) eye surgery in certain patient populations.

Product Classification:

Class II

Date Initiated: September 16, 2013
Date Posted: October 30, 2013
Recall Number: Z-0047-2014
Event ID: 66304
Reason for Recall:

This corrective action will address an identified failure of the lamp and illumination system within the laser unit caused by an unexpected failure of a thyristor component on the AC Power Distribution Printed Circuit Board (PCB).

Status: Terminated
Product Quantity: 834 units in total
Code Information:

All units of these models: STAR S2 Excimer Laser System; STAR S3 ActiveTrak; STAR S4 with Variable Spot Scanning; and STAR S41R; There are no expiration dates for the STAR Excimer Laser Systems.

Distribution Pattern:

Worldwide distribution: US (nationwide) including Puerto Rico and countries of: Asia Pacific, Canada, Europe, Africa, Middle East, Japan, and Latin America.

Voluntary or Mandated:

Voluntary: Firm initiated