Abbott Medical: Medical Device Recall in 2024 - (Recall #: Z-0493-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Abbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, Sterile
Product Classification:
Class II
Date Initiated: October 17, 2024
Date Posted: November 27, 2024
Recall Number: Z-0493-2025
Event ID: 95627
Reason for Recall:
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Status: Ongoing
Product Quantity: 3 OUS
Code Information:
UDI-DI (GTIN): 5415067045775, Serial Numbers: 20150441, 20156804, & 20170535
Distribution Pattern:
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Voluntary or Mandated:
Voluntary: Firm initiated