Abbott Medical: Medical Device Recall in 2024 - (Recall #: Z-2429-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Abbott Infinity 5, Implantable Pulse Generator, REF: 6660, SterileEO, Rx Only

Product Classification:

Class II

Date Initiated: May 22, 2024
Date Posted: July 31, 2024
Recall Number: Z-2429-2024
Event ID: 94796
Reason for Recall:

Their is a potential that the implantable pulse generator replacement indicator and end of service date may be shorter than indicated in the product labeling potentially resulting in loss of therapy.

Status: Ongoing
Product Quantity: 5,900 units
Code Information:

All Serial numbers/UDI:05415067030016/05415067020246

Distribution Pattern:

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, WA, WI, WY and the countries of Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, Ecuador, Estonia, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Netherlands, Norway, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Voluntary or Mandated:

Voluntary: Firm initiated