Abbott Point Of Care Inc.: Medical Device Recall in 2021 - (Recall #: Z-0881-2021)
See the recall detail below. You can also see other recalls from the same firm in 2021.
Abbott i-STAT EG7+ cartridge-IVD. Intended for the quantitative measurement of sodium, potassium, ionized calcium, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood. List Number: 03P76-25
Class II
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
Nationwide
Voluntary: Firm initiated