Abbott Vascular Inc: Medical Device Recall in 2025 - (Recall #: Z-0731-2025)
See the recall detail below. You can also see other recalls from the same firm in 2025.
20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation catheters, atherectomy devices, stent delivery systems, intravascular ultrasound devices).
Class II
Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
UDI-DI: 08717648013614, Lots: 60538750, 60538751, 60539113, 60545329
US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
Voluntary: Firm initiated