Abbott: Medical Device Recall in 2019 - (Recall #: Z-1220-2019)
See the recall detail below. You can also see other recalls from the same firm in 2019.
St. Jude Medical Confirm Rx Insertable Cardiac Monitor, REF DM3500, Sterile. Product Usage: The St. Jude Medical Confirm Rx insertable cardiac monitor (ICM) is designed to detect arrhythmias and wirelessly transmit data to the Merlin.net Patient Care Network (PCN). It is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.
Class II
The device is unable to pair with the mobile app due to the device incorrectly determining the certificate has expired.
Serial numbers 9225191, 9224501, 9225179, 9225193, 9225207, 9242209, 9257638, 9257646, 9257656, 9257659, 9224508, 9224490, 9224503, 9224525, 9224529, 9225190, 9225203, 9225205, 9225208, 9225211, 9225214, 9225851, 9225855, 9225858, 9225859, 9225867, 9225868, 9242199, 9242219, 9242225, 9242231, and 9256057.
US Nationwide Distribution in the states of AZ, CA, CO, ID, IL, LA, MA, MI, MO, NE, NH, NJ, NY, TN, TX, and WI.
Voluntary: Firm initiated