Abbott: Medical Device Recall in 2020 - (Recall #: Z-1700-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
Merlin PCS 3650 Software Upgrade Kit, REF 3330 St. Jude Medical The MerlinTM Patient Care System (MerlinTM PCS) Model 3650 is a portable, dedicated programming system designed to interrogate, program, display data, and test St. Jude MedicalTM implantable devices and leads.
Class II
Internal testing of software identified software anomaly scenarios where if non-applicable shock energy value is selected, the device will continue to use the previously programmed energy level and the user will not be alerted that the selected value was not retained.
Software Version: 24.6.1 Model: 3650, GTIN: 05414734500661 Model: 3330, GTIN: 05414734509725 For affected serial number of programmers see Attachment B.1 Programmers For models of ICDs interrogated by affected programmers see Attachment B.2 Devices.
U.S. AK, Al, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. OUS: None - the firm stated that there are no OUS consignees associated to ICDs that were interrogated from January 22, 2020 to February 6, 2020.
Voluntary: Firm initiated