Abiomed, Inc.: Medical Device Recall in 2024 - (Recall #: Z-0648-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Abiomed Oscor Introducer Kit, 14Fr x Long Individual (0052-0017)- Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3015

Product Classification:

Class II

Date Initiated: November 7, 2023
Date Posted: January 17, 2024
Recall Number: Z-0648-2024
Event ID: 93501
Reason for Recall:

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Status: Ongoing
Product Quantity: 3 single units
Code Information:

UDI-DI: N/A

Distribution Pattern:

Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK

Voluntary or Mandated:

Voluntary: Firm initiated