Accelerate Diagnostics Inc: Medical Device Recall in 2019 - (Recall #: Z-1065-2019)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2019.

Data Source: FDA.
Product Description:

Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.

Product Classification:

Class II

Date Initiated: March 13, 2017
Date Posted: April 3, 2019
Recall Number: Z-1065-2019
Event ID: 82216
Reason for Recall:

Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.

Status: Terminated
Product Quantity: 81
Code Information:

REF: 10401008. Serial Numbers: 1324, 1325, 1431, 1432, 1084, 1087, 1367, 1362,1363,1615, 1616, 1358, 1359, 1453, 1354, 1357, 1599, 1176, 1178, 1353, 1356, 1576, 1577, 1578, 1322, 1323, 1454, 1594, 1598, 1288, 1290, 1688, 1690, 1691, 1694, 1798, 1799, 1800, 1801, 1135, 1151, 1326, 1327, 1477, 1478, 1479, 1480, 1619, 1620, 1617, 1618, 1489, 1501, 1687, 1573, 1574, 1144, 1145, 1247, 1249 ,1264, 1277, 1265, 1277, 1278, 1381, 1150, 1286, 1148, 1149, 1254, 1260, 1397, 1542, 1458, 1517, 1231, 1238, 1282, 1284, 1389

Distribution Pattern:

Worldwide Distribution - US Nationwide North America and Middle East US: TX, IL, TN, GA, CA, AR, MD, VA, Washington DC, IA, PA, NY, WA, and MO. OUS (Foreign): UK, Spain, Italy, and Germany

Voluntary or Mandated:

Voluntary: Firm initiated