Acclarent, Inc.: Medical Device Recall in 2020 - (Recall #: Z-0825-2020)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2020.
Data Source: FDA.
Product Description:
TruDi NAV Suction, TDNS070Z, 70 Degree Tip Angle (01) 10705031245884
Product Classification:
Class II
Date Initiated: March 6, 2019
Date Posted: January 29, 2020
Recall Number: Z-0825-2020
Event ID: 84398
Reason for Recall:
The firm became aware that three (3) lots of the NAV Suction Instruments contain calibration issues causing incorrect tip orientation display in the system.
Status: Terminated
Product Quantity:
Code Information:
Lot # 1901080, 1901177, 1901219 The TruDi NAV Suction Instruments have a shelf life of 30 reprocessing cycles.
Distribution Pattern:
US: WI, IL, NE, CA, OH, NY OUS: None
Voluntary or Mandated:
Voluntary: Firm initiated