Ad-Tech Medical Instrument Corporation: Medical Device Recall in 2024 - (Recall #: Z-0220-2025)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
AD-TECH Spencer Probe Depth Electrode, REF SD06R-AP58X-000 6-Contact 1.12mm (A style) Depth Electrode
Product Classification:
Class II
Date Initiated: September 19, 2024
Date Posted: October 30, 2024
Recall Number: Z-0220-2025
Event ID: 95414
Reason for Recall:
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Status: Ongoing
Product Quantity: 4 units
Code Information:
UDI/DI 00841823107589, Lot Number 169260 858859860, 169258 858859860
Distribution Pattern:
US Nationwide distribution in the states of CA, OH.
Voluntary or Mandated:
Voluntary: Firm initiated