Advanced Sterilization Products: Medical Device Recall in 2013 - (Recall #: Z-0044-2014)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2013.

Data Source: FDA.
Product Description:

STERRAD 100NX Cassettes The STERRAD Sterilization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Product Classification:

Class II

Date Initiated: September 18, 2013
Date Posted: October 30, 2013
Recall Number: Z-0044-2014
Event ID: 66317
Reason for Recall:

Advanced Sterilization Products (ASP) is recalling the three lots (13F049, 13F050, 13F051) STERRAD 100NX Cassettes because ASP has determined that the leak indicator on the cassettes is effective for only 12 months of the cassette's labeled 15-month shelf-life.

Status: Terminated
Product Quantity: 2,995 units
Code Information:

Lot# 13F050, 13F051, 13F049

Distribution Pattern:

Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated