Aesculap Implant Systems LLC: Medical Device Recall in 2017 - (Recall #: Z-2087-2017)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2017.

Data Source: FDA.
Product Description:

GP161R; ELAN 4 1-RING DIAMOND BURR COARSE D2.0; GP162R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2.0; GP163R; ELAN 4 1-RING DIAMOND BURR COARSE D2,3; GP164R; ELAN 4 1-RING DIAMOND BURR+ COARSE D2,3; GP165R; ELAN 4 1-RING DIAMOND BURR COARSE D3,0; GP166R; ELAN 4 1-RING DIAMOND BURR+ COARSE D3,0; GP167R; ELAN 4 1-RING DIAMOND BURR++ COARSE D3,0; GP168R; ELAN 4 1-RING DIAMOND BURR COARSE D4,0; GP169R; ELAN 4 1-RING DIAMOND BURR COARSE D5,0; GP170R; ELAN 4 1-RING DIAMOND BURR COARSE D6,0; GP173R; ELAN 4 1-RING DIAMOND BURR X-COARSE D3,0; GP174R; ELAN 4 1-RING DIAMOND BURR X-COARSE D4,0; GP175R; ELAN 4 1-RING DIAMOND BURR X-COARSE D5,0; GP176R; ELAN 4 1-RING DIAMOND BURR X-COARSE D6,0; GP177R; ELAN 4 1-RING DIAMOND BURR X-COARSE D7,0; GP328R; ELAN 4 2-RING DIAMOND BURR X-COARSE D4.0; GP329R; ELAN 4 2-RING DIAMOND BURR X-COARSE D5.0; GP330R; ELAN 4 2-RING DIAMOND BURR X-COARSE D6.0; Product Usage: The ELAN 4 motor system and the Hi-Line XS handpiece are intended for high speed cutting, sawing, drilling and manipulation of soft tissue and bone in the fields of Spine, ENT, Neuro, and Maxillofacial Surgery.

Product Classification:

Class II

Date Initiated: March 28, 2017
Date Posted: May 24, 2017
Recall Number: Z-2087-2017
Event ID: 76988
Reason for Recall:

Diamond particulates may break off the burr and fall into the surgical site. If the particles are not detected and are not removed from the surgical site a foreign body reaction may occur. These particulates could also lead to mechanically induced soft tissue damage.

Status: Terminated
Product Quantity: Total of all burrs - 608 units
Code Information:

Model numbers COARSE: GP161R ; GP162R ; GP163R ; GP164R ; GP165R ; GP166R ; GP167R ; GP168R ; GP169R ; GP170R ; Model numbers X-Course: GP173R ; GP174R ; GP175R ; GP176R ; GP177R ; GP328R ; GP329R ; GP330R ;

Distribution Pattern:

US Nationwide distribution - CA, IL, NY, PA, TN, TX

Voluntary or Mandated:

Voluntary: Firm initiated