Aesculap, Inc.: Medical Device Recall in 2015 - (Recall #: Z-0448-2016)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2015.
Data Source: FDA.
Product Description:
Tray Set containing multiple instruments and may contain the flexible bone awl, flexible drill and/or flexible screw driver in varying amounts depending on order. Tray set is not sealed. These instruments are intended for use during the implantations of the A-Space SIBD and Arcadius XP L systems. Tray catalog number is: Tray designated as ST0485-037.
Product Classification:
Class II
Date Initiated: November 10, 2015
Date Posted: December 30, 2015
Recall Number: Z-0448-2016
Event ID: 72646
Reason for Recall:
The flexible shaft of the instruments may break during orthopedic spinal surgery. There is a risk of a delay in surgery or that a fragment from the shaft could be left in a patient.
Status: Terminated
Product Quantity: unknown
Code Information:
All Lots
Distribution Pattern:
Worldwide Distribution: US (nationwide) and country of: Canada.
Voluntary or Mandated:
Voluntary: Firm initiated