Aesculap, Inc.: Medical Device Recall in 2015 - (Recall #: Z-1012-2015)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2015.

Data Source: FDA.
Product Description:

Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Product Classification:

Class II

Date Initiated: December 17, 2014
Date Posted: February 4, 2015
Recall Number: Z-1012-2015
Event ID: 70086
Reason for Recall:

Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

Status: Terminated
Product Quantity: 772
Code Information:

material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T

Distribution Pattern:

Worldwide Distribution - US Nationwide

Voluntary or Mandated:

Voluntary: Firm initiated