Agiliti Health - Ellis: Medical Device Recall in 2026 - (Recall #: Z-1472-2026)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2026.

Data Source: FDA.
Product Description:

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Product Classification:

Class II

Date Initiated: January 21, 2026
Date Posted: March 4, 2026
Recall Number: Z-1472-2026
Event ID: 98332
Reason for Recall:

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Status: Ongoing
Product Quantity: 4286
Code Information:

UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated