Aircraft Medical Limited: Medical Device Recall in 2014 - (Recall #: Z-0494-2015)
See the recall detail below. You can also see other recalls from the same firm in 2014.
McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes
Class II
The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.
Product Code: 350-017-000. Lot Numbers and Expiration Dates: Lot 13121202, Expires 2016/12/11; Lot 13122301, Expires 2016/12/25; Lot 13042401, Expires 2017/5/5; Lot 14030501, Expires 2017/3/16; Lot 14031814, Expires 2017/3/26; Lot 14061616, Expires 2017/6/25.
Worldwide Distribution - US Nationwide in the state of GA and the countries of Austria, China, Denmark, Finland, France, Hong Kong, Israel, Italy, Netherlands, Poland, Saudi Arabia, Singapore, South Korea, Taiwan, and Turkey.
Voluntary: Firm initiated