Product Description:

McGrath MAC2 Video Laryngoscope Blades Product supplied in box of 50 items, each individually peel-pouched and EtO sterilized. Product Usage: Single Use, sterile video laryngoscope blade for use with McGRATH MAC Video Laryngoscopes

Product Classification:

Class II

Date Initiated: November 4, 2014
Date Posted: December 10, 2014
Recall Number: Z-0494-2015
Event ID: 69805
Reason for Recall:

The blade retaining clip may be difficult to engage and thereby increase the possibility of a blade becoming disconnected during intubation, specifically when the device is being retracted from the patients mouth.

Status: Terminated
Product Quantity: 40,900 blades (US); 16,000 blades (Foreign)
Code Information:

Product Code: 350-017-000. Lot Numbers and Expiration Dates: Lot 13121202, Expires 2016/12/11; Lot 13122301, Expires 2016/12/25; Lot 13042401, Expires 2017/5/5; Lot 14030501, Expires 2017/3/16; Lot 14031814, Expires 2017/3/26; Lot 14061616, Expires 2017/6/25.

Distribution Pattern:

Worldwide Distribution - US Nationwide in the state of GA and the countries of Austria, China, Denmark, Finland, France, Hong Kong, Israel, Italy, Netherlands, Poland, Saudi Arabia, Singapore, South Korea, Taiwan, and Turkey.

Voluntary or Mandated:

Voluntary: Firm initiated