Aizu Olympus Co., Ltd.: Medical Device Recall in 2023 - (Recall #: Z-0336-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2023.

Data Source: FDA.
Product Description:

Hysterovideoscope-For endoscopic diagnosis within the uterus Model Number: HYF-V.

Product Classification:

Class II

Date Initiated: September 25, 2023
Date Posted: November 29, 2023
Recall Number: Z-0336-2024
Event ID: 93222
Reason for Recall:

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Status: Ongoing
Product Quantity: 27 units
Code Information:

Model Number (UDI-DI): HYF-V (04953170340185).

Distribution Pattern:

Worldwide - US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated