Aizu Olympus Co., Ltd.: Medical Device Recall in 2023 - (Recall #: Z-0340-2024)
See the recall detail below. You can also see other recalls from the same firm in 2023.
Mobile airway scope-diagnosis and observation to access airway anatomy, endotracheal/endobronchial intubation and management. Model Numbers: MAF-DM2, MAF-GM, MAF-GM2, MAF-TM, MAF-TM2.
Class II
Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection
Model Numbers: MAF-DM2 (04953170407260), MAF-GM (04953170340369), MAF-GM2 (04953170407208), MAF-TM (04953170288630), MAF-TM2 (04953170407147).
Worldwide - US Nationwide distribution.
Voluntary: Firm initiated