Alcon Research, LLC: Medical Device Recall in 2023 - (Recall #: Z-0099-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Constellation Vision System, REF: 8065751150
Product Classification:
Class II
Date Initiated: August 18, 2023
Date Posted: October 18, 2023
Recall Number: Z-0099-2024
Event ID: 93010
Reason for Recall:
Ophthalmic microsurgical vision systems have a new Printed Circuit Board (PCB) design that could lead to higher voltage to a PCB component, that could cause component failure and loss of pressure control for the use of cutters, scissors, and forceps, which could lead to an increased risk of microbial infection, tissue damage and progressive visual impairment.
Status: Ongoing
Product Quantity: 3
Code Information:
UDI-DI: 00380657511501, Serial Numbers: 1403028701X, 1202865201X, 0802983301X
Distribution Pattern:
US: GA, IL, PR
Voluntary or Mandated:
Voluntary: Firm initiated