Alcon Research, LTD.: Medical Device Recall in 2018 - (Recall #: Z-0185-2019)
See the recall detail below. You can also see other recalls from the same firm in 2018.
CyPass System 241-S, CYPASS MICRO-STENT, CYPASS LOADED, CYPASS APPLIER, REF CATALOG #: 8065754003 For ophthalmic use.
Class I
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
GTIN: 00380657540037 All lot/serial numbers
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
Voluntary: Firm initiated