Alcor Scientific, Inc.: Medical Device Recall in 2022 - (Recall #: Z-1751-2022)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2022.
Data Source: FDA.
Product Description:
iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002
Product Classification:
Class II
Date Initiated: July 20, 2022
Date Posted: September 21, 2022
Recall Number: Z-1751-2022
Event ID: 90762
Reason for Recall:
Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
Status: Ongoing
Product Quantity: 38 units
Code Information:
Lot 4/22
Distribution Pattern:
US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
Voluntary or Mandated:
Voluntary: Firm initiated