Alere San Diego, Inc.: Medical Device Recall in 2012 - (Recall #: Z-1929-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Alere Triage Cardiac Panel PN 97000HS Used for the quantitative determination of creatine kinase MD (CK--MB), myoglobin and troponin I in EDTA anticoagulated whole blood or plasma specimens.

Product Classification:

Class I

Date Initiated: May 22, 2012
Date Posted: July 18, 2012
Recall Number: Z-1929-2012
Event ID: 61927
Reason for Recall:

Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.

Status: Terminated
Product Quantity: 6,044 kits
Code Information:

Phase 1: K50514B, W49753B, W49754B, W49756B, W49757B, W49758B, W49760B, W49761B, W49762B, W49764B, W49767B, W49768B, W49769B, W49770B, W50608B, W50613B, W50621B, W50623B, W50625B, W50632B, W50634B, W50637B, W50639B, W50641B, W50642B, W51114B, and W51119B. Phase 2: W49766B, W50605B, W50627B, W50628B, and W50629B.

Distribution Pattern:

Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated