Alere San Diego, Inc.: Medical Device Recall in 2012 - (Recall #: Z-1929-2012)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Alere Triage Cardiac Panel PN 97000HS Used for the quantitative determination of creatine kinase MD (CK--MB), myoglobin and troponin I in EDTA anticoagulated whole blood or plasma specimens.
Class I
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Phase 1: K50514B, W49753B, W49754B, W49756B, W49757B, W49758B, W49760B, W49761B, W49762B, W49764B, W49767B, W49768B, W49769B, W49770B, W50608B, W50613B, W50621B, W50623B, W50625B, W50632B, W50634B, W50637B, W50639B, W50641B, W50642B, W51114B, and W51119B. Phase 2: W49766B, W50605B, W50627B, W50628B, and W50629B.
Nationwide Distribution
Voluntary: Firm initiated