Alere San Diego, Inc.: Medical Device Recall in 2012 - (Recall #: Z-1931-2012)
See the recall detail below. You can also see other recalls from the same firm in 2012.
Alere Triage BNP PN 98000XR Intended to be used as an aid in the diagnosis and assessment of severity of congestive heart failure (also referred to as heart failure).
Class I
Certain lots of the affected products may have significantly decreased precision relative to the package insert, which could result in an increased frequency of false positive or false negative results. These false positive and false negative results are unpredictable within lots and may not be detected by Quality Control testing.
Phase 1: K50671B, W49486B, W49487B, W49489B, W49498B, W49500B, W49873B, W49890B, W49895B, W49896B, W49934B, W49947B, W49948B, W49950B, W49962B, W49968B, W50257B, W50278B, W50279B, W50282B, W50287B, W50294B, and W50313B. Phase 2: K50665B, W49855B, W49862B, W49865B, W49887B, W49892B, W49893B, W49894B, W49900B, W49933B, W49939B, W49953B, W49958B, W49960B, W49961B, W50252B, W50268B, W50273B, W50274B, W50290B, W50292B, W50295B, W50297B, W50298B, W50304B, W50355B, W50357B, and W50375B.
Nationwide Distribution
Voluntary: Firm initiated