Alere San Diego, Inc.: Medical Device Recall in 2012 - (Recall #: Z-2039-2012)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2012.

Data Source: FDA.
Product Description:

Alere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescence immunoassay to be used with the Alere Triage Meters for the qualitative determination of the presence of drug and/or the major metabolites above the threshold concentrations of up to 10 distinct drug classes, including assays for acetaminophen/paracetamol, amphetamines, methamphetamines, barbiturates, benzodiazepines, cocaine, methadone, opiates, phencyclidine, THC and tricyclic antidepressants in urine. The acetaminophen/paracetamol assay will yield positive results when acetaminophen/paracetamol is ingested at or above therapeutic doses. This test provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS), Liquid Chromatography/ Mass Spectrometry/ Mass Spectrometry (LC/MS/MS) and High Performance Liquid Chromatography (HPLC) are common confirmatory methods. A quantitative serum acetaminophen/paracetamol measurement is the common confirmatory method for preliminary positive acetaminophen/paracetamol results.

Product Classification:

Class II

Date Initiated: May 14, 2012
Date Posted: July 25, 2012
Recall Number: Z-2039-2012
Event ID: 62293
Reason for Recall:

A recall was initiated because Alere San Diego has confirmed that the Triage TOX Drug Screen may have significantly decreased precision and accuracy than claimed. An increased frequency of false positive and/or false negative results has been observed for Triage TOX Drug Screen.

Status: Terminated
Product Quantity: 9,997 kits
Code Information:

Lot/Unit Codes: W50216B, W50222B, W50245B, W50250B, W50809B, W50810B, W50812B, W50813B, W50816B, W50817B, W50818B, W50819B, W50820B, W50821B, W50822B, W50823B, W50826B, W50829B, W50830B, W50831B, W50832B, W50837B, W50838B, W50842B, W50843B, W50844B, W50846B, W50848B, W50851B, W50862B, W50866B, W50869B, W50870B, W50871B, W50872B, W50873B.

Distribution Pattern:

US Nationwide Distribution

Voluntary or Mandated:

Voluntary: Firm initiated