Alere San Diego, Inc.: Medical Device Recall in 2016 - (Recall #: Z-2778-2016)
See the recall detail below. You can also see other recalls from the same firm in 2016.
LDX Optical Check Cassettes (OCC) Model Number: 10-228 and 14-535 Cassette in box with foam insert. The cassette is used to check the optics system of the LDX analyzer. Product Usage: The Optics Check Cassette is used to check the optics system of the Alere Cholestech LDX¿ Analyzer. It is run once a day before patient samples are tested and after the analyzer has been moved or serviced.
Class III
Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.
Model Number: 10-228 and 14-535 Lot Number: 332994
Worldwide Distribution - U.S Nationwide in the states of: TN, MA, GA, OK, VA, FL, TN, MO, AL, TX, IL, NY, VT, RI, KY, WI, AL, NC, LA, SD, TX, VA, MI, MN, IN, IA, CA, AR, PA, HI, WV, KS, CO, CT, UT and the countries of Japan, England, Canada
Voluntary: Firm initiated