Alphatec Spine, Inc.: Medical Device Recall in 2014 - (Recall #: Z-2092-2014)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2014.
Data Source: FDA.
Product Description:
Solanas 2.1mm FIXED DEPTH DRILL, 10mm, Part No. 63928-10. Component of the SOLANAS III Posterior Stabilization System, a spinal fixation system intended to improve stability of the cervical and thoracolumbar area of the spine (C1-T3).
Product Classification:
Class II
Date Initiated: June 26, 2014
Date Posted: July 30, 2014
Recall Number: Z-2092-2014
Event ID: 68667
Reason for Recall:
Alphatec Spine is recalling the 2.1mm Fixed Depth Drill 10mm because they received a report that the instrument malfunctioned and golden stop slid from its fixed position during use.
Status: Terminated
Product Quantity: 14 units total (12 units in US)
Code Information:
Lot No. 7117902
Distribution Pattern:
Nationwide Distribution including CT, NC, AL, WV, IL, AR, and NY.
Voluntary or Mandated:
Voluntary: Firm initiated