Ameditech Inc: Medical Device Recall in 2020 - (Recall #: Z-3029-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Alere iScreen Single Drug Detection (AMT Item Number/ AMT Description): X11-IS1 BUP-DX/ "Alere iScreen" Dx Single Dip Card Single Drug Detection: Buprenorphine (BUP)", X11-IS1 BAR-DX /"Alere iScreen" Dx Single Dip Card Single Drug Detection: Barbiturate (BAR)", X11-IS1 OXY-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Oxycodone (OXY)", X11-IS1 BZO-DX / "Alere iScreen" Dx Single Dip Card Single Drug Detection: Benzodiazepine (BZO300)", X11-IS1 MET-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methamphetamine (MET)", X11-IS1 AMP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Amphetamine (AMP)", X11-IS1 MDM-DX / "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Ecstasy (MDMA)", X11-IS1 MTD-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Methadone (MTD)", X11-IS1 MOP-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Morphine (MOP)", X11-IS1 THC-DX/"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Marijuana (THC)", X11-IS1 TCA-DX/ "Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Tricyclic Antidrepessants (TCA)", X11-IS1 PCP-DX /"Alere iScreen¿ Dx Single Dip Card Single Drug Detection: Phencyclidine (PCP)",

Product Classification:

Class II

Date Initiated: August 18, 2020
Date Posted: October 7, 2020
Recall Number: Z-3029-2020
Event ID: 86335
Reason for Recall:

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Status: Terminated
Product Quantity: 303,100 units
Code Information:

ALL Lots Within Expiry

Distribution Pattern:

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Voluntary or Mandated:

Voluntary: Firm initiated