Ameditech Inc: Medical Device Recall in 2020 - (Recall #: Z-3036-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."
Class II
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
ALL Lots Within Expiry
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Voluntary: Firm initiated