Ameditech Inc: Medical Device Recall in 2020 - (Recall #: Z-3039-2020)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2020.

Data Source: FDA.
Product Description:

Noble 1 Step+ Cup (AMT Item Number/ AMT Description): NBCA-13A-B /"Noble 1 Step + Cup" THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; w/ Cr, Ni, pH, Bl, S.G. Finished Devices", NBCA-14A-PP/"Noble 1 Step + Cup" (THC50/COC150)+(OPI300/MET500)+ (AMP500/BZO300)+(BAR300/MTD300)+ (BOPG10/TCA1000)+(MDMA500/OXY100)+ (PCP25/PPX300) w/ Cr, Ni, pH, Bl, S.G."

Product Classification:

Class II

Date Initiated: August 18, 2020
Date Posted: October 7, 2020
Recall Number: Z-3039-2020
Event ID: 86335
Reason for Recall:

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Status: Terminated
Product Quantity: 13,700 units
Code Information:

ALL Lots Within Expiry

Distribution Pattern:

US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain

Voluntary or Mandated:

Voluntary: Firm initiated