Ameditech Inc: Medical Device Recall in 2020 - (Recall #: Z-3045-2020)
See the recall detail below. You can also see other recalls from the same firm in 2020.
ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300", HT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Adulteration Screen for CR - NI - PH - BL - S.G.", PT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/OPI300/THC50" DT13A/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100 w/ Cr, Ni, pH, BI, S.G.", DT14A "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, Bl, S.G. Finished Devices", DT12A/"ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, BI, S.G.", PT12A/ "ToxCup¿ Drug Screen Cup COC300/MET500/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", PT11A/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", HT30/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET1000/PCP25/THC50/ OPI2000"
Class II
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
ALL Lots Within Expiry
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Voluntary: Firm initiated