American Contract Systems, Inc.: Medical Device Recall in 2023 - (Recall #: Z-0182-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2023.
Data Source: FDA.
Product Description:
Cath Cardiac Pack, REF IHCC03V
Product Classification:
Class II
Date Initiated: September 1, 2023
Date Posted: November 1, 2023
Recall Number: Z-0182-2024
Event ID: 93086
Reason for Recall:
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Status: Ongoing
Product Quantity: 1260 units
Code Information:
UDI/DI 00191072189431, Lot Numbers: 724221, 679221, 945231
Distribution Pattern:
US Nationwide distribution in the states of FL, IA.
Voluntary or Mandated:
Voluntary: Firm initiated