American Contract Systems, Inc.: Medical Device Recall in 2024 - (Recall #: Z-0824-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
Custom procedural convenience kits and trays, general & plastic surgery, labeled as: a) Breast Plastics Pack, kit number BBPL20D; b) MINOR PLASTIC, kit number WPMP16H
Class II
During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
a) BBPL20D, Lot Codes: 862221, exp. 5/18/2023; 688221, exp. 5/31/2023; 756221, exp. 9/1/2023; 782221, exp. 8/6/2023; b) WPMP16H,Lot Codes: 904231, exp. 4/6/2024; 966231, exp. 2/3/2024; 650221, exp. 12/16/2023; 708221, exp. 10/19/2023; 805221, exp. 7/14/2023; 888221, exp. 4/22/2023; 867221, exp. 5/13/2023
US Nationwide distribution.
Voluntary: Firm initiated