American Contract Systems, Inc.: Medical Device Recall in 2024 - (Recall #: Z-0826-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

Custom procedural convenience kits and trays, general hospital use, labeled as: a) Baby PICC Kit, kit number UIBP90J; b) Baby PICC Kit, kit number UIBP90K; c) PORT PACK (PS 058330), kit number UIPO82AB; d) Pediatric PICC Kit, kit number UIPP76K;

Product Classification:

Class II

Date Initiated: November 29, 2023
Date Posted: February 7, 2024
Recall Number: Z-0826-2024
Event ID: 93656
Reason for Recall:

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Status: Ongoing
Product Quantity: 244 units
Code Information:

a) UIBP90J, UDI/DI 00191072117038, Lot Numbers: 996231, exp. 1/4/2024; b) UIBP90K, UDI/DI 00191072189356, Lot Numbers: 911231, exp. 7/26/2023; c) UIPO82AB, UDI/DI 00191072172006, Lot Numbers: 727221, exp. 9/30/2023; 680221, exp. 11/16/2023; d) UIPP76K, UDI/DI 00191072142849, Lot Numbers: 637221, exp. 12/29/2023; 976231, exp. 1/24/2024; 898231, exp. 4/12/2024

Distribution Pattern:

US Nationwide distribution.

Voluntary or Mandated:

Voluntary: Firm initiated