American Contract Systems, Inc.: Medical Device Recall in 2024 - (Recall #: Z-2521-2024)

Updated on March 25, 2026.

See the recall detail below. You can also see other recalls from the same firm in 2024.

Data Source: FDA.
Product Description:

American Contract Systems Extremity convenience kit, American Contract Systems Hand Pack convenience kit, and American Contract Systems Total Knee Pack convenience kit

Product Classification:

Class II

Date Initiated: May 6, 2024
Date Posted: August 14, 2024
Recall Number: Z-2521-2024
Event ID: 94908
Reason for Recall:

Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility

Status: Ongoing
Product Quantity: 15 kits
Code Information:

(1) Model No ALEX17A; UDI-DI 00191072201546, Lot 983241;Exp. Date 1/17/2025; (2) Model No AKHD97A; UDI-DI 00191072201126; Lot 988241;Exp. Date 1/12/2025; (3) Model No CETJ13O; UDI-DI 00191072202017; Lot 989241;Exp. Date 1/11/2025.

Distribution Pattern:

US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.

Voluntary or Mandated:

Voluntary: Firm initiated