American Contract Systems, Inc.: Medical Device Recall in 2024 - (Recall #: Z-2521-2024)
See the recall detail below. You can also see other recalls from the same firm in 2024.
American Contract Systems Extremity convenience kit, American Contract Systems Hand Pack convenience kit, and American Contract Systems Total Knee Pack convenience kit
Class II
Product released with insufficient aeration time impacting the kits sterilized on the same cycle or in the same aeration chamber, which may have been compromised in sterility
(1) Model No ALEX17A; UDI-DI 00191072201546, Lot 983241;Exp. Date 1/17/2025; (2) Model No AKHD97A; UDI-DI 00191072201126; Lot 988241;Exp. Date 1/12/2025; (3) Model No CETJ13O; UDI-DI 00191072202017; Lot 989241;Exp. Date 1/11/2025.
US Nationwide distribution in the states of CA, KS, ME, MN, MO, NE, and SD.
Voluntary: Firm initiated