American Contract Systems, Inc.: Medical Device Recall in 2024 - (Recall #: Z-3229-2024)
Updated on March 25, 2026.
See the recall detail below. You can also see other recalls from the same firm in 2024.
Data Source: FDA.
Product Description:
Arthroscopy Pack, MSKA45A; Medical convenience kit
Product Classification:
Class II
Date Initiated: August 7, 2024
Date Posted: October 2, 2024
Recall Number: Z-3229-2024
Event ID: 95261
Reason for Recall:
ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Status: Ongoing
Product Quantity: 48 units
Code Information:
UDI-DI: 00191072200938; Lot: 2405153; Exp: 05/15/2025
Distribution Pattern:
US Nationwide distribution in the states of MO, MN, MA, OH, NE.
Voluntary or Mandated:
Voluntary: Firm initiated