Product Description:
SD-210U-15 ELECTROSURGICAL SNARE, SD-210U-15 - Product Usage: Used with endoscopes to electrosurgically resect tissue within the digestive tract.
Product Classification:
Class II
Date Initiated: January 4, 2021
Date Posted: February 17, 2021
Recall Number: Z-0985-2021
Event ID: 87074
Reason for Recall:
Inspection confirmed a small percentage of sterile packages with inadequate package seal (partial heat seal). Patients may develop infections if non-sterile equipment is used.
Status: Terminated
Product Quantity: 7,400,000 (global); 260,395 (US)
Code Information:
05K, 07K, 09K, 92K, 93K, 95K, 96K, 97K, 98K, 99K, 9XK, 9YK, 97V, 98V, 99V, 9XV, 03V, 04V, 05V, 06V, 07V, 9YV
Distribution Pattern:
US Nationwide distribution.
Voluntary or Mandated: